YSI 200
|
Temperature
|
Range : -6 to +46¡ÆC
|
Resolution : 0.1¡ÆC
|
Accuracy : ¡¾0.3¡ÆC ¡¾1 digit
|
DO % Air Saturation
|
Range : 0 to 200%
|
Resolution : 0.1% air saturation
|
Accuracy : ¡¾2% of the reading or ¡¾0.2 air saturation, whichever is greater
|
DO ppm (mg/L)
|
Range : 0 to 20 ppm
|
Resolution : 0.01 ppm
|
Accuracy : ¡¾2% of the reading or ¡¾0.2 ppm, whichever is greater
|
Accessories
(order cable separately)
|
200-4 : 4 meter probe and cable assembly
|
200-10 : 10 meter probe and cable assembly
|
280 : DO measurement carrying case (molded plastic with foam insert)
|
5908 : Membrane Kit, 1.25 PE membrane with probe solution
|
480 : Soft, black carrying case with shoulder strap
|
Additional Technical Specifications
|
Operating range
|
Temperature : 0¡ÆC to 50¡ÆC
|
Relative Humidity : up to 95%
|
Water resistance
|
IP-65 water resistant, splash resistant
|
Size
|
70mm width x 186mm length (2.8in. x 7.3in)
|
Weight with battery
|
350g (0.75lb)
|
ATC Probe
|
Thermistor, 10§Ú/25¡ÆC
|
Battery
|
One 9volt included with purchase
|
Salinity compensation
|
0.0 to 40.0 ppt
|
Pressure compensation
|
600 to 1100 millibar (450 to 825 mmHg)
|
Calibration back-up
|
Yes
|
Audio feedback
|
Yes, all keys
|
|
|