low price levitra

levitra coupon
KANOMAX,6621P/6631P,Air Velocity Meter,dz¼Ó°è,ÇÁ¸°ÅÍ ³»ÀåÇü dz¼Ó°è,dz¼Ó ÃøÁ¤,¿­¼±½Ä ÃøÁ¤±â,Hot Wire Air Velocity Meter">

 ÀÀ¿ë ±â°èº°(¹°)

 ÀÀ¿ë ±â°èº°(°ø±â)

 ±â»ó °üÃøÀåÄ¡

 ±âŸ °üÃøÀåÄ¡

 Á¦ÀÛ ½ÇÇèÀåÄ¡

 

 

 

  ÇöÀçÀ§Ä¡ : HOME > Á¦Ç°Á¤º¸ > ÀÀ¿ë ±â°èº° > dz¼Ó/dz·® ÃøÁ¤ ÀåÄ¡ > 6621P/6631P

 

          6621P/6631P (ÇÁ¸°ÅÍ ³»ÀåÇü Ç³¼Ó°è)

 

  Maintenance and examination of air conditioning system.

 

  Monitoring of air flow in clean room or clean booth.

 

  Measurement of air velocity and static pressure in the duct.

 

 

 

 

 

 

 

 

  Equipped with useful calculation functions.

      (Max., Min. and average value/amount of air/display by a graph)

 

  Digital output terminal to transfer the data to PC.

 

  Telescopic probe enable to use for many purpose.

 

 

 

 

 

   Model 6621-P / 6631-P

Measuring Function

 Air velocity, air temperature and static pressure

 (static pressure by 6631 only)

Measuring Object

 0¡­100¡É clean airflow

Measuring Range

 Air velocity         : 0.1¡­50m/s     Air temperature  : 0¡­100¡É

 Static pressure  : 0¡­¡¾5kPa

Measuring Accuracy

and indication resolving

power

 

Range 

Accuracy

Indication Resolving

Power

Air Velocity

0.1¡­5m/s

¡¾0.1m/s

0.01m/s

5¡­10m/s

¡¾0.2m/s

10¡­25m/s

¡¾0.5m/s

0.1m/s

25¡­50m/s

¡¾1.0m/s

Air temperature

0¡­100¡É

¡¾0.5¡É

0.1¡É

Static Pressure

¡¾0¡­1kPa

¡¾0.03kPa

¡¾0.01kPa

¡¾1¡­2kPa

¡¾0.05kPa

¡¾2¡­5kPa

¡¾0.1kPa

Temperature

 compensation accuracy

 ¡¾5% FS in a temperature range of 5¡­80¡É

Response

 Air velocity       : Approx. 1sec (90% at air velocity of 1m/s)

 Air temperature : Approx. 10sec (90% at air velocity of 1m/s)

 Static pressure : Less than 1sec (90% at less than 3m of

 connecting tube)

Display

 Digital (simultaneous display of air velocity and temperature or static

 pressure)

Analog output terminal

 DC 0¡­1V (output impedance 47§Ù) ... Model 6621-PA, 6631-PA only

 Simultaneous 2ch output for air velocity and temperature

 (static pressure changeover)

Digital output terminal

 RS-232C (1,200 or 4,800 bps)

Printer

 Thermal type (built-in)

Memory

 Max. 999 data (in case of 1-page/measurement)

Power supply

 Ni-Cd Battery, AC Adaptor is only for recharge (DC12.5V, 450mA)

Operating temperature

 5¡­40¡É

Dimensions

 Main body : 318 x 222 x 48mm

 Probe : 190¡­480(flexible) dimension of extension rod ¨ª10¡­14mm

Weight

 Main body : Approx. 1.9§¸,  Porbe : Approx. 200g (curl cord include)

Components

 Main Unit(1), Probe(1), Curl Cord(1), Ni-Cd Battery & AC Adaptor(1),

 Printer Paper(1)

 Thermal Printer(1), Pressure Tube 2m(1) ... Model 6631 only,

 Operation   Manual(1)

 

 